 Invasive suction machine
Hits:7285    AddTime:2020-04-20

Basic Requirements:

multiple breathing mode, apply to ICU, emergency department, Operating Room, and home ward use; 

advanced synchronous function, can effectively avoid man-machine confrontation, can be used without trauma / trauma. 

microcomputer control, constant volume, pressure trigger multi-function suction machine, digital display mode. 

Functional Parameters:

ventilation mode: interval positive pressure ventilation (IPPV) , interval positive pressure ventilation + Sigh (IPPV + Sigh) , synchronous interval positive pressure ventilation (SIPPV) , synchronous interval directive ventilation + continuous flow (SIMV) , Sigh Sigh Sigh Sigh Sigh Sigh, PEEP, MANUAL (MANUAL)

Tidal Volume (TV) : At least 100-1200ml; 

Inspiratory Time: 0.5-3s;

Inhalation Frequency: 1-60BPM; 

oxygen concentration (O2%) : 45-100% , continuously adjustable; 

Synchronous Trigger Sensitivity:-0.5-1.0KPA; 

Sigh (Sigh) : Given every 100 times or as needed;

PEEP: 0.5-2 KPA continuously adjustable, digital display 

Airway pressure:-2 ~ 10 KPA 

safety pressure ≤7.2 KPA 

Monitor display parameters:

tidal volume (minute ventilation) 

breathing frequency

Airway pressure (peak pressure, positive end pressure, trigger pressure) 

alarm setting value,

Status Indication (power, inhale, breath, pressure column digital indication)

Warning Instructions: 

Upper limit of pressure (5-60HPA) 

Lower limit of pressure (2-40hpa) ,

Upper limit of ventilation (10-60L / Min) 

Lower limit of ventilation (4-20L / Min) ,

overfrequency breathing (10-110bpm) ,

ASPHYXIA

Incorrect setting of parameters,

Power failure,

alarm silencing

security protection: 

Airway pressure protection,

accidental shutdown or power failure protection

parameter setting error protection;

Basic configuration: 

by the host, 

Mobile Trolley,

mechanical arm,

adjustable temperature humidifier

the Suction Tube Group (suction valve can be disassembled high-temperature high-pressure disinfection, silicone suction tube and sampling tube) and other components. 

Others: 

WITH ISO9001 AND ISO13485 Quality System Certification;

【Refresh】【Favorites】【Print】 【Close】